Datum der Veröffentlichung: 2022/12/12
Typ: Kostenlose Testversion (30 Tage)
Plattformen: Windows 10 with .Net Framework 6
Minimale Plattformanforderungen:
- Intel i5 2,5 GHz
- 8 GB RAM
- 512 MB Grafikkarte
- Netzwerkadapter mit 1GBPS
Unterstützte Sprachen: Englisch, Deutsch, Französisch, Spanisch, Italienisch, Portugiesisch (Brasilien), Japanisch, Koreanisch, Chinesisch (vereinfacht), Russian, Czech, Swedish
*CRITICAL*
When applying emissivity to an individual Region Of Interest (ROI), other ROIs may also have this emissivity applied if set to follow frame emissivity (indicated by ‘-‘ in the emissivity column of the ROI table). This issue affects all versions between 2.1.0 and 2.1.5.
It is recommended that all customers with version 2.1.* upgrade to V2.1.6. The current license will still be valid.
[It is also possible fix this issue by applying an explicit emissivity to each ROI in the ROI settings]
Versionshinweise:
- Ability to change ROI value reported by GTV TCP command to Max, Mean or Min
- Improved layout for Isotherm editor.
Bug Fixes
-
OPC settings dialog incorrectly reports 'Port In Use' fixed
-
SCF TCP command auto-appends file extension if not supplied
-
ROI name change is now propagated to the ROI Group editor
-
Histogram percent values now shown with '%' in Simple Value
-
Fixes internal TagID creation issue on playback control.